Emergency Procedures
From BrainImagingCenter
Download Emergency Procedures as .doc file (24 KB)
Emergency Stop Buttons
- Control Room (near door)
- Scan Suite Wall (near door)
- Hallway wall to equipment room
- System Power Panel (equipment room back wall)
Emergency Procedures
- Fire:
- Emergency Stop
- Remove subject from area
- Alert campus security (dial 3333)
- Extinguish (2 extinguishers, waiting room/equipment room)
- *** CLOSE SCANNER SUITE DOOR ***
- Medical:
- Stop Scan
- Remove subject from scanner suite
- Alert campus security (dial 3333)
- Tend to the subjects needs
- *** CLOSE SCANNER SUITE DOOR ***
- Quench:
- Turn on fans
- Press quench button
- Evacuate scanner and scanner suite immediately
- Alert campus security (dial 3333)
- Tend to subjects needs
- *** CLOSE SCANNER SUITE DOOR ***
Equipment Problems
- No Chirping:
- Call Terry Sackett
- Call Jim Haxby
- Humidity Alarm:
- Let it go: FO&M will come and check it out
- Scanner Technical Issues:
- Most issues can be cleared up by rebooting (must restart system, not just log out of software)
- Call Terry Sackett
- Call Jim Haxby